LED Power Solutions

powered by Current Automation

OWA-90U series

90W Single Output Moistureproof Adaptor

 

Application:

Indoor LED lighting
LED Decorative Lighting
LED Office Lighting

Features

 Universal AC input/Full range
 Built in active PFC function
 High efficiency up to 91%
 Protections: Short circuit/ Overload / Over Voltage / Over Temperature
 Class II power unit, no FG
 Fanless design, cooling by free air convection
 No load power consumption < 0.15W
 Energy efficiency Level VI
 Comply with EISA 2007 DoE and NRCan
 UL LISTED, suitable for dry and damp locations
 5 year warranty

OWA-90U series

MODELOWA-90U-12 OWA-90U-15OWA-90U-20
OUTPUTDC VOLTAGE12V15V20V
CONSTANT CURRENT REGION7.2 ~ 12V 9 ~ 15V12 ~ 20V
RATED CURRENT7.5A6A4.5A
RATED POWER90W90W90W
RIPPLE & NOISE (max.) 150mVp-p150mVp-p150mVp-p
VOLTAGE TOLERANCE±4.0%±4.0%±4.0%
LINE REGULATION±0.5%±0.5%±0.5%
LOAD REGULATION±1.5%±1.0%±0.5%
SETUP TIME500ms, 80ms at 95% load 115VAC / 230VAC500ms, 80ms at 95% load 115VAC / 230VAC500ms, 80ms at 95% load 115VAC / 230VAC
HOLD UP TIME (Typ.)16ms at full load 115VAC / 230VAC16ms at full load 115VAC / 230VAC16ms at full load 115VAC / 230VAC
INPUTVOLTAGE RANGE90 ~ 264VAC 127 ~ 370VDC90 ~ 264VAC 127 ~ 370VDC90 ~ 264VAC 127 ~ 370VDC
FREQUENCY RANGE47 ~ 63Hz47 ~ 63Hz47 ~ 63Hz
POWER FACTOR (Typ.)PF>0.98/115VAC, PF>0.96/230VAC at full loadPF>0.98/115VAC, PF>0.96/230VAC at full loadPF>0.98/115VAC, PF>0.96/230VAC at full load
TOTALHARMONIC DISTORTIONTHD< 20% when output loading>= 60% at 115VAC/230VACTHD< 20% when output loading>= 60% at 115VAC/230VACTHD< 20% when output loading>= 60% at 115VAC/230VAC
EFFICIENCY (Typ.) - 115VAC88%89%89%
EFFICIENCY (Typ.) - 230VAC89%90%90%
AC CURRENT0.95A/ 115VAC 0.5A/ 230VAC0.95A/ 115VAC 0.5A/ 230VAC0.95A/ 115VAC 0.5A/ 230VAC
INRUSH CURRENT(Typ.) - 30ACOLD START 30A(twidth=550 s measured at 50% Ipeak) at 115VACCOLD START 30A(twidth=550 s measured at 50% Ipeak) at 115VACCOLD START 30A(twidth=550 s measured at 50% Ipeak) at 115VAC
INRUSH CURRENT(Typ.) - 60ACOLD START 60A(twidth=550 s measured at 50% Ipeak) at 230VACCOLD START 60A(twidth=550 s measured at 50% Ipeak) at 230VACCOLD START 60A(twidth=550 s measured at 50% Ipeak) at 230VAC
MAX. No. of PSUs on 16A
CIRCUIT BREAKER
3 units (circuit breaker of type B) / 6 units (circuit breaker of type C) at 230VAC3 units (circuit breaker of type B) / 6 units (circuit breaker of type C) at 230VAC3 units (circuit breaker of type B) / 6 units (circuit breaker of type C) at 230VAC
LEAKAGE CURRENT<0.125mA/ 120VAC <0.25mA/ 240VAC<0.125mA/ 120VAC <0.25mA/ 240VAC<0.125mA/ 120VAC <0.25mA/ 240VAC
Specs Sheet
untitled-3Download PDF
untitled-3Download PDF
untitled-3Download PDF

OWA-90U series

MODELOWA-90U-24OWA-90U-30OWA-90U-36
OUTPUTDC VOLTAGE24V30V36V
CONSTANT CURRENT REGION14.4 ~ 24V18 ~ 30V 21.6 ~ 36V
RATED CURRENT3.75A3A 2.5A
RATED POWER90W90W90W
RIPPLE & NOISE (max.) 150mVp-p200mVp-p 200mVp-p
VOLTAGE TOLERANCE±3.0%±3.0% ±2.0%
LINE REGULATION±0.5%±0.5%±0.5%
LOAD REGULATION±0.5% ±0.5% ±0.5%
SETUP TIME500ms, 80ms at 95% load 115VAC / 230VAC500ms, 80ms at 95% load 115VAC / 230VAC500ms, 80ms at 95% load 115VAC / 230VAC
HOLD UP TIME (Typ.)16ms at full load 115VAC / 230VAC16ms at full load 115VAC / 230VAC16ms at full load 115VAC / 230VAC
INPUTVOLTAGE RANGE90 ~ 264VAC 127 ~ 370VDC90 ~ 264VAC 127 ~ 370VDC90 ~ 264VAC 127 ~ 370VDC
FREQUENCY RANGE47 ~ 63Hz47 ~ 63Hz47 ~ 63Hz
POWER FACTOR (Typ.)PF>0.98/115VAC, PF>0.96/230VAC at full loadPF>0.98/115VAC, PF>0.96/230VAC at full loadPF>0.98/115VAC, PF>0.96/230VAC at full load
TOTALHARMONIC DISTORTIONTHD< 20% when output loading>= 60% at 115VAC/230VACTHD< 20% when output loading>= 60% at 115VAC/230VACTHD< 20% when output loading>= 60% at 115VAC/230VAC
EFFICIENCY (Typ.) - 115VAC89%89%89.5%
EFFICIENCY (Typ.) - 230VAC90%90%91%
AC CURRENT0.95A/ 115VAC 0.5A/ 230VAC0.95A/ 115VAC 0.5A/ 230VAC0.95A/ 115VAC 0.5A/ 230VAC
INRUSH CURRENT(Typ.) - 30ACOLD START 30A(twidth=550 s measured at 50% Ipeak) at 115VACCOLD START 30A(twidth=550 s measured at 50% Ipeak) at 115VACCOLD START 30A(twidth=550 s measured at 50% Ipeak) at 115VAC
INRUSH CURRENT(Typ.) - 60ACOLD START 60A(twidth=550 s measured at 50% Ipeak) at 230VACCOLD START 60A(twidth=550 s measured at 50% Ipeak) at 230VACCOLD START 60A(twidth=550 s measured at 50% Ipeak) at 230VAC
MAX. No. of PSUs on 16A
CIRCUIT BREAKER
3 units (circuit breaker of type B) / 6 units (circuit breaker of type C) at 230VAC3 units (circuit breaker of type B) / 6 units (circuit breaker of type C) at 230VAC3 units (circuit breaker of type B) / 6 units (circuit breaker of type C) at 230VAC
LEAKAGE CURRENT<0.125mA/ 120VAC <0.25mA/ 240VAC<0.125mA/ 120VAC <0.25mA/ 240VAC<0.125mA/ 120VAC <0.25mA/ 240VAC
Specs Sheet
untitled-3Download PDF
untitled-3Download PDF
untitled-3Download PDF

OWA-90U series

MODELOWA-90U-42OWA-90U-48OWA-90U-54
OUTPUTDC VOLTAGE42V48V54V
CONSTANT CURRENT REGION25.2 ~ 42V28.8 ~ 48V32.4 ~ 54V
RATED CURRENT2.15A1.88A1.67A
RATED POWER90.3W90.24W90.18W
RIPPLE & NOISE (max.) 250mVp-p 250mVp-p 350mVp-p
VOLTAGE TOLERANCE±1.0% ±1.0% ±1.0%
LINE REGULATION±0.5%±0.5%±0.5%
LOAD REGULATION±0.5% ±0.5% ±0.5%
SETUP TIME500ms, 80ms at 95% load 115VAC / 230VAC500ms, 80ms at 95% load 115VAC / 230VAC500ms, 80ms at 95% load 115VAC / 230VAC
HOLD UP TIME (Typ.)16ms at full load 115VAC / 230VAC16ms at full load 115VAC / 230VAC16ms at full load 115VAC / 230VAC
INPUTVOLTAGE RANGE90 ~ 264VAC 127 ~ 370VDC90 ~ 264VAC 127 ~ 370VDC90 ~ 264VAC 127 ~ 370VDC
FREQUENCY RANGE47 ~ 63Hz47 ~ 63Hz47 ~ 63Hz
POWER FACTOR (Typ.)PF>0.98/115VAC, PF>0.96/230VAC at full loadPF>0.98/115VAC, PF>0.96/230VAC at full loadPF>0.98/115VAC, PF>0.96/230VAC at full load
TOTALHARMONIC DISTORTIONTHD< 20% when output loading>= 60% at 115VAC/230VACTHD< 20% when output loading>= 60% at 115VAC/230VACTHD< 20% when output loading>= 60% at 115VAC/230VAC
EFFICIENCY (Typ.) - 115VAC89.5%89.5%89.5%
EFFICIENCY (Typ.) - 230VAC91%91%91%
AC CURRENT0.95A/ 115VAC 0.5A/ 230VAC0.95A/ 115VAC 0.5A/ 230VAC0.95A/ 115VAC 0.5A/ 230VAC
INRUSH CURRENT(Typ.) - 30ACOLD START 30A(twidth=550 s measured at 50% Ipeak) at 115VACCOLD START 30A(twidth=550 s measured at 50% Ipeak) at 115VACCOLD START 30A(twidth=550 s measured at 50% Ipeak) at 115VAC
INRUSH CURRENT(Typ.) - 60ACOLD START 60A(twidth=550 s measured at 50% Ipeak) at 230VACCOLD START 60A(twidth=550 s measured at 50% Ipeak) at 230VACCOLD START 60A(twidth=550 s measured at 50% Ipeak) at 230VAC
MAX. No. of PSUs on 16A
CIRCUIT BREAKER
3 units (circuit breaker of type B) / 6 units (circuit breaker of type C) at 230VAC3 units (circuit breaker of type B) / 6 units (circuit breaker of type C) at 230VAC3 units (circuit breaker of type B) / 6 units (circuit breaker of type C) at 230VAC
LEAKAGE CURRENT<0.125mA/ 120VAC <0.25mA/ 240VAC<0.125mA/ 120VAC <0.25mA/ 240VAC<0.125mA/ 120VAC <0.25mA/ 240VAC
Specs Sheet
untitled-3Download PDF
untitled-3Download PDF
untitled-3Download PDF